pH 600 Series Handheld Waterproof pH Meter

These waterproof handheld meters feature a pH range up to -2.00 to 19.99 pH and resolutions up to 0.001 pH. Combined with accuracies up to ±0.001 pH, these meters provide a new benchamark in our pH meter line.

Store up to 500 data sets in the meter's memory. Time-and-date-stamping meets Good Laboratory Practice (GLP) standards.

The pH 600 series of meters offers the latest in calibration capabilities, including up to 6 calibration points, a user-settable calibration due alarm, and up to 15 buffer options with auto buffer recognition of USA, NIST, DIN, and PWB standards.In addition, the pH 620 meter is capable of measuring dissolved ion concentrations through the use of ion selective electrodes (ISE).

The infrared IrDA wireless technology makes PC downloading more convenient and easier. Dual power up allows you to use batteries while out in the field or the optional AC power adapter while in the lab.

As always, our waterproof meters are IP67-rated for both dust and water intrusion, even with no probes attached!

Order pH probes separately.


 
  • Expanded range, resolution, and accuracy, increased memory, and advanced communications!
  • No more excuses for missing data! Out-of-date or unperformed calibrations are now things of the past!
  • Don't waste time looking for those cables!
  • Use in dirty and wet environments.


Download pH600 Series Product Sheet (.pdf, 142KB)

SPECIFICATIONS

 
pH600 Handheld Meter
pH610 Handheld Meter
pH620 Handheld Meter
Range pH
-2.00 to 19.99 pH
-2.000 to 19.999 pH
-2.000 to 19.999 pH
mV
±2000 mV
±2000 mV
±2000 mV
ISE
-
-
0.001 to 19,900 ppm, molar, or mg/L
Temperature
-10 to 110°C (14 to 230°F)
-10 to 110°C (14 to 230°F)
-10 to 110°C (14 to 230°F)
Resolution pH
0.1/0.01 pH
0.1/0.01/0.001 pH
0.1/0.01/0.001 pH
mV
1/0.1 mV
1/0.1 mV
1/0.1 mV
ISE
-
-
2 or 3 digits
Temperature
0.1°C (0.1 to 1°F)
0.1°C (0.1 to 1°F)
0.1°C (0.1 to 1°F)
Accuracy pH
±0.01 pH
±0.002 pH
±0.002 pH
mV
±0.2/2.0 mV
±0.2/2.0 mV
±0.2/2.0 mV
ISE
-
-
0.5% FS (monovalent)
1% FS (divalent)
Temperature
±0.5°C (0.9°F)
±0.5°C (0.9°F)
±0.5°C (0.9°F)
Temperature Compensation
Automatic or manual, 0 to 100°C
Calibration (pH)
Up to 6 points
Buffer Recognition

USA: 1.68, 4.01, 7.00, 10.01, 12.45
NIST: 1.68, 4.01, 6.86, 9.18, 12.45
DIN: 1.09, 3.06, 4.65, 6.79, 9.23, 12.75
PWB: 4.10, 6.97

Data Logging
500 data sets with GLP
Output
IRDA/RS-232C bi-directional
Display
Dot-matrix LCD with backlighting
Power
Four AA Batteries (included)
Universal AC Adapter (optional)
Low Battery Indicator
Yes
Dimensions
3-1/4"W x 7-1/4"H x 2-1/4"D